No signal on Western blot, what to do next?-分析方法-资讯-生物在线

No signal on Western blot, what to do next?

作者:Agrisera AB 2020-10-21T00:00 (访问量:1435)

 No signal on Western blot, what to do next?

Western blot, one of most commonly used laboratory techniques for protein work, is quite complex and many factors can influence its outcome. In case of problems with the detection of a target protein, the following can be done:
Use a cellular fraction/organ in which the protein of interest is expressed. Many antibodies to proteins of low expression are not going to detect a signal in a total cell extract. 
Increase the protein load to 20-50 ug/well
Blocking: 5-10 % non-fat milk 1h/RT or ON/4°C
Affinity purified primary antibody may produce a signal not obtained in case of serum or total IgG fraction. 
Recommended primary antibody starting dilution: 
1: 500-1:1000/ON/4°C incubation
If a target is of low expression, the most sensitive detection methods should be applied: Chemiluminescence (ECL) or fluorescence is preferred over chromogenic detection (ALP-conjugated secondary antibodies)
Overexpose, to make sure that the band of interest does not require longer time to become visualized, and adjust your protocol accordingly. 

 
Western blot optimisation - before and after


Left: before optimization of western blot conditions
Right: after optimization of western blot conditions

For catalog antibodies, we recommend checking the product information sheet for suggested experimental conditions, including sample type, protein load and more. 

You are always welcome to contact Agrisera Technical Support and we will be happy to help. 
Agrisera AB 商家主页

地 址: 21 Vännäs, SWEDEN, Europe

联系人: Dr. Joanna Porankiewicz-Asplund

电 话: 0046704137107

传 真:

Email:61382170@qq.com

相关咨询
ADVERTISEMENT